Abstract
Terahertz (THz) time-domain spectroscopy has been investigated for assessment of the hydration levels in the cornea, intraocular pressure, and changes in corneal topography. Previous efforts at THz imaging of the cornea have employed off-axis parabolic mirrors to achieve normal incidence along the spherical surface. However, this comes at the cost of an asymmetric field-of-view (FOV) and a long scan time because it requires raster-scanning of the collimated beam across the large mirror diameter. This paper proposes a solution by designing a pair of aspheric lenses that can provide a larger symmetric spherical FOV (9.6 mm) and reduce the scan time by two orders of magnitude using a novel beam-steering approach. A hyperbolic-elliptical lens was designed and optimized to achieve normal incidence and phase-front matching between the focused THz beam and the target curvature. The lenses were machined from a slab of high-density polyethylene and characterized in comparison to ray-tracing simulations by imaging several targets of similar sizes to the cornea. Our experimental results showed excellent agreement in the increased symmetric FOV and confirmed the reduction in scan time to about 3-4 seconds. In the future, this lens design process can be extended for imaging the sclera of the eye and other curved biological surfaces, such as the nose and fingers.
Funder
Stony Brook University
National Institute of General Medical Sciences
Subject
Atomic and Molecular Physics, and Optics
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献