Abstract
This work presents a portable optical meter for noncontact thickness measurement. The
device shines a focused laser light on a thin and transparent sample,
resulting in an interference between light reflecting from the top and
from the bottom surface, and the interfering pattern is recorded by a
linear sensor array before data analysis with an Arduino
microcontroller. The device produced accurate thickness values from
glass cover slips and transparent plastic sheets within a fraction of
a second per measurement. Additionally, the sample’s refractive index
is not required a priori. Therefore, it
has a high potential to be of use in real-time quality control in
transparent thick-film coating and manufacturing.