Abstract
Spectral micro-CT imaging with direct-detection energy discriminating photon counting detectors having small pixel size (< 100×100 µm2) is mainly hampered by: i) the limited energy resolution of the imaging device due to charge sharing effects and ii) the unavoidable noise amplification in the images resulting from basis material decomposition. In this work, we present a cone-beam micro-CT setup that includes a CdTe photon counting detector implementing a charge summing hardware solution to correct for the charge-sharing issue and an innovative image processing pipeline based on accurate modeling of the spectral response of the imaging system, an improved basis material decomposition (BMD) algorithm named minimum-residual BMD (MR-BMD), and self-supervised deep convolutional denoising. Experimental tomographic projections having a pixel size of 45×45 µm2 of a plastinated mouse sample including I, Ba, and Gd small cuvettes were acquired. Results demonstrate the capability of the combined hardware and software tools to sharply discriminate even between materials having their K-Edge separated by a few keV, such as e.g., I and Ba. By evaluating the quality of the reconstructed decomposed images (water, bone, I, Ba, and Gd), the quantitative performances of the spectral system are here assessed and discussed.
Funder
Commissione Scientifica Nazionale 5, Instituto Nazionale di Fisica Nucleare
Subject
Atomic and Molecular Physics, and Optics
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献