Affiliation:
1. Technical University of Denmark
Abstract
We demonstrate a high dynamic range (DR) Fourier-transform-based terahertz (THz) spectrometer by combining a THz photomultiplier tube (PMT) with a metasurface and a conventional Michelson interferometer. Because the THz-PMT response depends on the incident electric-field strength following the Fowler–Nordheim equation, we can directly obtain an electric field interferogram without any synchronized optical probe pulse in contrast to conventional THz-time-domain-spectroscopy (THz-TDS). The DR of the corresponding power spectrum using the proposed method was 4.6 × 105 without the use of a lock-in amplifier. The complex refractive index of a quartz glass plate obtained using the proposed method was in good agreement with the results of conventional THz-TDS.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献