Radiometric cross-calibration of Sentinel-2B MSI with HY-1C SCS based on the near simultaneous imaging of common ground targets

Author:

Xu Heyu1ORCID,Huang Wenxin1,Si Xiaolong1,Song Qingjun2ORCID,Li Xin1,Weiwei Xu1,Ma Yue3,Zhang Liming1

Affiliation:

1. Chinese Academy of Sciences

2. Ministry of Natural Resources of the People’s Republic of China

3. Beijing Institute of Space Mechanics & Electricity

Abstract

To simplify the cross-calibration process and improve calibration frequency and accuracy, this paper proposes a cross-calibration method for the multispectral remote sensor Multi-Spectral Instrument (MSI) carried by Sentinel-2B using the hyperspectral remote sensor, that is, the satellite calibration spectrometer (SCS) carried by Hai Yang (HY)-1C, as the reference sensor and establishes the calibration process. Precise spectral response matching between SCS and MSI spectral channels is performed by the interpolation and iteration of hyperspectral data to eliminate the difference in band settings and significantly improve the accuracy of cross-calibration coefficients. The SNO-x inherited from the simultaneous nadir overpass (SNO) method is used as a prediction method to carry out cross-calibration imaging in mid- and low-latitude regions, which improves the cross-calibration frequency and broadens the dynamic range of calibration. The cross-calibration coefficients and offsets of MSI B1∼B7 and B8a were obtained by processing the earth observation images of the MSI and SCS on January 24, 2019. Then, the cross-calibration coefficients and offsets are applied to the ocean, farmland and other ground objects with different reflectance, and the reliability and accuracy of the cross-calibration results are evaluated with the Moderate-resolution Imaging Spectroradiometer (MODIS) carried by Terra as a reference. To improve the accuracy of the evaluation, the spectral band adjustment factor between the corresponding channels of MSI and MODIS is used to correct the measured reflectance of MODIS based on the satellite calibration coefficient. The reflectance directly obtained by processing the MSI image is used as the MSI-measured reflectance, the reflectance obtained based on the cross-calibration coefficient is used as the MSI-calculated reflectance, and the reflectance corrected by the spectral band adjustment factor (SBAF) is used as the MODIS-calculated reflectance. The results show that the mean root-mean-square relative error (RMSRE) between the MODIS-calculated reflectance and the MSI-calculated reflectance is 2.16% and that the mean RMSRE between the MODIS-calculated reflectance and the MSI-measured reflectance is 3.05%, indicating that the reflectance corrected based on calibration coefficients is closer to the MODIS-calculated reflectance. Finally, each uncertainty source in the cross-calibration is analyzed, and the comprehensive uncertainty is found to be 4.03%, indicating that SCS can be used as a reference for MSI cross-calibration.

Funder

Hefei Institutes of Physical Science, Chinese Academy of Sciences

Publisher

Optica Publishing Group

Subject

Atomic and Molecular Physics, and Optics

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3