Affiliation:
1. Hangzhou Dianzi University
Abstract
To realize the high sensitivity polarization sensitive optical coherence tomography (PS-OCT) imaging, a fiber-based full-range depth-encoded swept source PS-OCT (SS-PS-OCT) method is proposed. The two OCT images corresponding to the orthogonal polarized input light are located on the high sensitivity imaging region of the opposite sides relative to the zero optical path difference position. The full-range OCT images can be obtained by implementing the spatial phase modulation in the reference arm. The detection sensitivity of the system was measured experimentally to be 67 dB when the imaging depth approaching to 2 mm. The imaging of the biological tissue verifies that the proposed full-range depth-encoded SS-PS-OCT system has the higher detection sensitivity compared with the conventional depth encoded SS-PS-OCT system. Finally, we demonstrated the full-range high sensitivity phase retardation image of the bovine tendon and skin of human fingertip. The fiber-based full-range depth-encoded SS-PS-OCT method can realize the high sensitivity birefringence imaging in the medical diagnosis scenes with the requirements for long imaging range and high detection sensitivity.
Funder
National Natural Science Foundation of China