Affiliation:
1. Hefei Comprehensive National Science Center
2. Huazhong University of Science and Technology
Abstract
High-frequency (greater than 30 MHz) photoacoustic computed tomography (PACT) provides the opportunity to reveal finer details of biological tissues with high spatial resolution. To record photoacoustic signals above 30 MHz, sampling rates higher than 60 MHz are required according to the Nyquist sampling criterion. However, the highest sampling rates supported by existing PACT systems are typically within the range of 40–60 MHz. Herein, we propose a novel PACT imaging method based on sub-Nyquist sampling. The results of numerical simulation, phantom experiment, and in vivo experiment demonstrate that the proposed imaging method can achieve high-frequency PACT imaging with a relatively low sampling rate. An axial resolution of 22 μm is achieved with a 30-MHz transducer and a 41.67-MHz sampling rate. To the best of our knowledge, this is the highest axial resolution ever achieved in PACT based on a sampling rate of not greater than 60 MHz. This work is expected to provide a practical way for high-frequency PACT imaging with limited sampling rates.
Funder
National Key Research and Development Program of China
National Natural Science Foundation of China
Anhui Provincial Department of Science and Technology
University of Science and Technology of China
Institute of Artificial Intelligence at Hefei Comprehensive National Science Center