Abstract
Real-time transrectal ultrasound (TRUS) image guidance during robot-assisted laparoscopic radical prostatectomy has the potential to enhance surgery outcomes. Whether conventional or photoacoustic TRUS is used, the robotic system and the TRUS must be registered to each other. Accurate registration can be performed using photoacoustic (PA markers). However, this requires a manual search by an assistant [IEEE Robot. Autom. Lett 8, 1287 (2023).10.1109/LRA.2022.3191788]. This paper introduces the first automatic search for PA markers using a transrectal ultrasound robot. This effectively reduces the challenges associated with the da Vinci-TRUS registration. This paper investigated the performance of three search algorithms in simulation and experiment: Weighted Average (WA), Golden Section Search (GSS), and Ternary Search (TS). For validation, a surgical prostate scenario was mimicked and various ex vivo tissues were tested. As a result, the WA algorithm can achieve 0.53°±0.30° average error after 9 data acquisitions, while the TS and GSS algorithm can achieve 0.29∘±0.31∘ and 0.48°±0.32° average errors after 28 data acquisitions.
Funder
Intuitive Surgical
Johns Hopkins University
Canadian Institutes of Health Research
National Institutes of Health
National Science Foundation
Subject
Atomic and Molecular Physics, and Optics,Biotechnology