Affiliation:
1. MPI f. Neorobiol. of Behavior – caesar
Abstract
Laser scanning microscopy requires beam steering through relay and focusing optics at sub-micron precision. In light-weight mobile systems, such as head mounted multiphoton microscopes, distortion and imaging plane curvature management is unpractical due to the complexity of required optic compensation. Thus, the resulting scan pattern limits anatomical fidelity and decreases analysis algorithm efficiency. Here, we present a technique that reconstructs the three-dimensional scan path only requiring translation of a simple fluorescent test probe. Our method is applicable to any type of scanning instrument with sectioning capabilities without prior assumptions regarding origin of imaging deviations. Further, we demonstrate that the obtained scan pattern allows analysis of these errors, and allows to restore anatomical accuracy relevant for complementary methods such as motion correction, further enhancing spatial registration and feature extraction.
Subject
Atomic and Molecular Physics, and Optics,Biotechnology
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献