Affiliation:
1. Tianjin Key Laboratory for Control Theory and Applications in Complicated Systems
Abstract
In this study, a high-precision rotation angle measurement method based on polarization self-mixing interference (SMI) is proposed. The higher signal-to-noise ratio SMI signal can be obtained by the differential processing of two polarized SMI signals with opposite phases. In order to reduce the influence of the speckle effect, the envelope signal is used to normalize the SMI signal. The fringe subdivision method is used to improve the accuracy of the rotation angle measurement. The experimental results show that the error of the rotation angle measurement is within ±0.5%, and the measurement range can reach up to 20°.
Funder
National Natural Science Foundation of China
Subject
Atomic and Molecular Physics, and Optics,Engineering (miscellaneous),Electrical and Electronic Engineering
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献