Abstract
We propose a simple, cost-effective method for synchronized phase contrast and fluorescence video acquisition in live samples. Counter-phased pulses of phase contrast illumination and fluorescence excitation light are synchronized with the exposure of the two fields of an interlaced camera sensor. This results in a video sequence in which each frame contains both exposure modes, each in half of its pixels. The method allows real-time acquisition and display of synchronized and spatially aligned phase contrast and fluorescence image sequences that can be separated by de-interlacing in two independent videos. The method can be implemented on any fluorescence microscope with a camera port without needing to modify the optical path.