Abstract
Photoacoustic imaging with optical resolution usually requires a single-pixel raster scan. An alternative approach based on illumination with patterns obtained from a Hadamard matrix, measurement of the generated ultrasound wave with a single detector, followed by a reconstruction known from computational ghost imaging is demonstrated here. Since many pixels on the object are illuminated at the same time, thereby contributing to the recorded signal, this approach gives a better contrast-to-noise ratio compared to the raster scan, as demonstrated in a phantom experiment. Furthermore, exploiting the temporal information for depth-resolved imaging is possible. The proposed method will be beneficial in situations where the radiant exposure of a sample is limited due to either safety precautions or the properties of the available light source.
Funder
Österreichische Forschungsförderungsgesellschaft
Subject
Atomic and Molecular Physics, and Optics
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献