Affiliation:
1. Universidad Nacional Autónoma de México
2. Universidad del Istmo
Abstract
To measure the shape of the fast corneal surface of the human eye, we propose the design and characterization of a compact corneal topographer using the capabilities of a smartphone. The performance evaluation of the compact corneal topographer includes the calculation and compensation of the distortion introduced by the smartphone lens used to acquire the images and the evaluation of a reference surface. To demonstrate the feasibility of our proposal, we performed surface topography measurements on some human corneas and compared the results with those obtained by a commercial corneal topographer. We showed that the results obtained with our algorithms were consistent with other methods that analyze the corneal surface.
Funder
Consejo Nacional de Ciencia y Tecnología
Dirección General de Asuntos del Personal Académico, Universidad Nacional Autónoma de México
Subject
Atomic and Molecular Physics, and Optics,Engineering (miscellaneous),Electrical and Electronic Engineering
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献