Affiliation:
1. The Third People’s Hospital of Shenzhen
Abstract
Regular and narrow-band RGB cameras are recently explored for contactless SpO2 monitoring. Regular RGB cameras with cross-band overlap provide a high signal-to-noise-ratio (SNR) in measuring the photoplethysmographic signals but possess high dependency on the spectra of incident light, whereas narrow-band RGB cameras have better spectral independence but lower SNR especially in dim lighting conditions, such as in the neonatal intensive care unit (NICU). This paper proposes a notch RGB camera based SpO2 measurement approach that uses an optical notch filter to attenuate the wavelengths of 580–605 nm of a regular RGB camera to improve the spectral independence while maintaining high SNR in signal measurement. The proposed setup was validated in the lab condition (e.g. dark chamber) against the existing solutions for visible-light based camera-SpO2 measurement and further verified in the NICU on preterm infants. The clinical trial conducted in the NICU with 22 preterm infants shows that the notch RGB camera can achieve a mean absolute error (MAE) less than 4% for SpO2 measurement. This is the first showcase of continuous monitoring of absolute camera-SpO2 values in the NICU.
Funder
National Key Research and Development Program of China
National Natural Science Foundation of China
Basic and Applied Basic Research Foundation of Guangdong Province
Shenzhen Science and Technology Innovation Program
Shenzhen Fundamental Research Program
Subject
Atomic and Molecular Physics, and Optics,Biotechnology
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献