Affiliation:
1. China University of Petroleum
Abstract
An optimization model for the parameters of the pressure-resistant lens of an optical particle counter, based on Gaussian optics theory, was established to increase the measurement accuracy of the counter for high-pressure natural gas. Comparing the experimental and calculated values of the calibrated model, when the pressure-resistant lens is displaced by 2 mm under atmospheric pressure, the relative error of the measured body deformation is 0.15%. When the air pressure varies in the range 0.10–5.09 MPa, the maximum relative error of optical measurement volume deformation with the change of refractive index is 0.13%, which shows that the model has high reliability and accuracy.
Funder
National Key Research and Development Program of China
Subject
Atomic and Molecular Physics, and Optics,Engineering (miscellaneous),Electrical and Electronic Engineering
Cited by
3 articles.