Validation of hemoglobin measurement in venous blood using HemoCue for the Ensanut 2022
-
Published:2023-11-13
Issue:6, nov-dic
Volume:65
Page:612-619
-
ISSN:1606-7916
-
Container-title:Salud Pública de México
-
language:
-
Short-container-title:Salud Publica Mex
Author:
De la Cruz-Góngora Vanessa,Méndez-Gómez Humarán Ignacio,Jaimes-Terán Socorro,Gaytán Miguel,Gómez-Acosta Luz María,Vizuet-Vega Norma Isela,Gaona-Pineda Elsa Berenice,Cuevas-Nasu Lucia,Romero Martín,Dary Omar,Shamah-Levy Teresa
Abstract
Objective. To validate hemoglobin (Hb) concentration in venous blood (VB) using the HemoCue 201+ in a subsample of children and women from the Mexican National Health and Nutrition Survey 2022. Materials and methods. Prior to field work 50 HemoCue 201+ devices were verified using venous blood. During the field work 57 children (aged 1-11) and 62 women (aged 12-49) donated 3 mL of VB each. Hb was measured in each device of HemoCue 201+ and in a hematologic autoanalyzer for the validation. Results. No significant bias was found in most of the devices. An adjustment criterion was used for 22 devices. Hb mean difference results were -0.049±0.578 g/dL in children and -0.098±0.628 g/dL in women. Conclusions. The HemoCue 201+ is a valid tool for estimating Hb concentration to produce reliable estimates of anemia prevalence when using venous blood.
Publisher
Instituto Nacional de Salud Publica
Subject
Public Health, Environmental and Occupational Health,Public Health, Environmental and Occupational Health
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献