Comparative dosimetric, setup margin, and treatment time analysis between ring gantry and C-Arm linear accelerators for VMAT-based craniospinal irradiation plans

Author:

Biswal Subhra S.12,Sarkar Biplab12,Goyal Monika2

Affiliation:

1. Department of Radiation Oncology, Apolo Multispeciality Hospitals, Kolkata, West Bengal

2. Depertment of Physics, GLA University, Mathura, Uttar Pradesh, India

Abstract

Purpose: The purpose of this study is to evaluate the dosimetric and treatment delivery characteristics of volumetric modulated arc therapy technique (VMAT)-based craniospinal axis irradiation (CSI) between ring gantry Halcyon (HAL) and C-arm based Novalis Tx (NTx) linear accelerator. Set-up margin and treatment delivery time for both machines were also taken into account. Materials and Methods: Fifteen patients, 4 females and 11 males treated between March 2019 and February 2022 within the age group 4–56 years simulated in the supine position and were planned for multiple isocentre VMAT technique in ring gantry Halcyon and C-Arm Novalis linear accelerator for 6FFF and 6 MV flatten beam energy. The number of isocenters was the same in both the machines, usually three for adult adolescent age group patients and two for pediatric patients. Total on-couch time and the patient positional shift were captured for each isocenter during each session of treatment. Margins were calculated using Herk’s formula of margin = 2.5Σ +0.7σ. Dosimetry, on-couch time, and set-up margin were compared between two competing arms. Results: Ninety-five percent of PTV coverage (P = 0.333), volume receiving 107% (P = 0.676), total MU (P = 0.818) in both the arms were comparable and statically insignificant. Low-dose spillage such as D20% (P = 0.212) and D50% (P = 0.008) was lesser in HAL comparable to NTx. CI and HI were statically insignificant. Out of 26 organs at risk (OAR), only 3 organs showed a statically significant dose difference. The mean and maximum setup margin in any linear direction was 0.45 and 0.53 cm for HAL and 0.37 and 0.56 cm for NTx and, variation was statistically insignificant (0.23 < P < 0.47). On-couch time was 4.0 ± 5.5 min lesser for HAL and the difference in on-couch time between the two arms was statistically different. Conclusion: Even though the majority of the delivery parameters such as gantry speed, dose rate, beam characteristic (flatten or unflatten), MLC width, and speed between the ring gantry HAL and C-arm NTx linear accelerators were distinctly different, they offered no or minimal difference in the dose distribution and in the setup margin. HAL gives a faster treatment time delivery, which could be crucial for some selective cases such as patients receiving treatment under general anesthesia.

Publisher

Medknow

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3