The feasibility of the pretreatment verification of 2D dose distributions in radiation therapy with small fields using the electronic portal imaging device

Author:

Tuncel Nina1,Mahdavi Seied Rabi2,Haghparast Mohammad3,Karakus Ismail1,Haghparast Abbas4,Nazari Vahideh5

Affiliation:

1. Radiation Oncology Department, School of Medicine, Akdeniz University, Antalya, Turkey

2. Department of Medical Physics, School of Medicine, Iran University of Medical Sciences, Tehran, Iran

3. Department of Radiology, Faculty of Paramedicine, Hormozgan University of Medical Sciences, Bandar Abbas, Iran

4. Department of Medical Physics, Faculty of Medicine, Kermanshah University of Medical Sciences, Kermanshah, Iran

5. Department of Medical Physics, Hamadan University of Medical Sciences, Hamadan, Iran

Abstract

Abstract Background: The present study aims to evaluate the performance of an Electronic portal imaging device (EPID) for measuring dosimetric parameters and for verification of dose in small photon fields. Material and Methods: In this study, the beam profiles were obtained using the amorphous silicon (a-Si) EPID for field sizes ranging from 1 × 1 to 10 × 10 cm2 at energies 6 and 18 mega-voltage (MV). For comparison, the dosimetric parameters, including penumbra widths and field sizes, were measured with the pinpoint, diode, and Semiflex dosimeters. Finally, Rando Phantom was used to compare the two-dimensional (2D) Dose distribution between EPID and Treatment Planning System (TPS). Results: In both 5 cm and 10 cm depths, there were large differences between the measured doses obtained from TPS, Pinpoint detector, and Farmer detector in 1 × 1 field size. The differences become negligible as the field sizes increase and from 3 × 3 field size to 10 × 10 field size, the maximum observed differences are 2 cGy and 2.4 cGy for 5 cm and 10 cm depths, respectively. The results indicate that the penumbra widths are smaller in the Gantry-Target (GT) direction compared to the Right-Left (RL) direction. The maximum difference (47.6%) was observed for EPID in the 10 × 10 field size, and the minimum difference (16.6%) was observed for TPS in the 1 × 1 field size. Finally, 2D dose distributions obtained by EPID and TPS exhibit excellent agreement. Conclusion: EPID is an excellent tool for the measurement of dosimetry parameters such as dose profiles, penumbra widths, field sizes, and pretreatment verification of 2D dose distributions, especially in small fields.

Publisher

Medknow

Subject

Radiology, Nuclear Medicine and imaging,Oncology,General Medicine

Reference13 articles.

1. Dose-response characteristics of an amorphous silicon EPID;Winkler;Med Phys,2005

2. EPID dosimetry for pretreatment quality assurance with two commercial systems;Bailey;J Appl Clin Med Phys,2012

3. Evaluation of Perkin elmer amorphous silicon electronic portal imaging device for small photon field dosimetry;Haghparast;J Biomed Phys Eng,2022

4. Transit dosimetry with an electronic portal imaging device (EPID) for 115 prostate cancer patients;Pasma;Int J Radiat Oncol,1999

5. A method for deconvolution of integrated electronic portal images to obtain incident fluence for dose reconstruction;Renner;J Appl Clin Med Phys,2005