Comparison and estimation of photoneutron dose produce between 10 MV flattened and unflattened beam in Elekta Versa HD™ medical linac

Author:

Khilafath Hajee Reyaz Ali Sahib1,Ganesan Bharanidharan1,Sekar Nandakumar1,Mohapatra Dinakrushna1,Vellingiri Jayaprakash2,Prakasarao Aruna1,Mahadevan Pramod2,Singaravelu Ganesan1

Affiliation:

1. Department of Medical Physics, College of Engineering, Anna University, Chennai, Tamil Nadu, India

2. Department of Radiation Oncology, VPS Lakeshore Hospital, Kochi, Kerala, India

Abstract

ABSTRACT Background: In a high-energy medical linear accelerator (linac), if the interaction of photon energy is higher than the neutron binding energy of high atomic material, it emits a neutron field through photonuclear (γ, n) reaction. Aim: The current study, evaluates the photoneutron dose equivalent (PNDE) produced between the 10 MV flattened and unflattened beams as a function of field sizes in the Elekta Versa HD™ linac. Materials and Methods: The PNDE produced from Versa linac was recorded along the patient plane using the bubble detector personal neutron dosimeter and from the measured PNDE values, the theoretical PNDE values were simulated for various field sizes using nonlinear least-squares curve-fitting as a function of a polynomial. The percentage of deviation (PoD) and Chi-square (χ2) tests were performed between the measured and simulated PNDE values to study the reliability and validity. Results: The results show that the mean PoD between the measured and simulated PNDE values for respective positions of a field size of FF beam was found to be -1.99% for 0.3×0.3, -4.39% for 5×5, -3.868% for 10×10, 0.590% for 15×15, 9.18% for 20×20, -4.133% for 25×25, and 0.467% for 30×30 cm2. Similarly, the mean PoD between the measured and simulated PNDE values for flattening filter-free (FFF) beam was found to be 1.36% for 0.3×0.3, -1.39% for 5×5, -5.38% for 10×10, 4.41% for 15×15, 3.84% for 20×20, 5.69% for 25×25, and -1.75% for 30×30 cm2. The maximum deviation between the measured and simulated PNDE values lies within the range ± 5%. Conclusions: From the study, it is observed that the FFF beam produces lesser neutron contamination than the FF beam.

Publisher

Medknow

Subject

Radiology, Nuclear Medicine and imaging,Oncology,General Medicine

Reference24 articles.

1. Cancer incidence and mortality worldwide:Sources, methods and major patterns in GLOBOCAN 2012;Ferlay;Int J Can,2015

2. Survival benefit of radiotherapy after surgery in de novo stage IV breast cancer:A population-based propensity-score matched analysis;Kim;Sci Rep,2019

3. The tumour response to induction chemotherapy has prognostic value for long-term survival outcomes after intensity-modulated radiation therapy in nasopharyngeal carcinoma;Peng;Sci Rep,2016

4. Neutron source in the Varian clinac 2100/2300C medical accelerator calculated by the EGS4 code;Mao;Health Phys,1997

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3