Jiménez-Puertas Sara1,Rodríguez Andrea González1,Cordero Sergio Lozares1,González Tomás González1,Chamarro Javier Díez1,Hernández Mónica Hernández1,Moreno Raquel Castro1,Casi Marta Sánchez1,Gazulla David Carlos Villa1,Martínez Almudena Gandía1,Bonel Arantxa Campos1,Valiño Maria del Mar Puertas1,Gómez José Antonio Font1

Affiliation:

1. Department of Medical Physics and Radiation Protection, University Hospital Miguel Servet, P.º de Isabel la Católica, 1-3, 50009 Zaragoza, Spain

Abstract

Purpose: This study aims to minimize monitor units (MUs) of intensity-modulated treatments in the Monaco treatment planning system while preserving plan quality by optimizing the “Minimum Segment Width” (MSW) and “Fluence Smoothing” parameters. Materials and Methods: We retrospectively analyzed 30 prostate, 30 gynecological, 15 breast cancer, 10 head and neck tumor, 11 radiosurgery, and 10 hypo-fractionated plans. Original prostate plans employed “Fluence Smoothing” = Off and were reoptimized with Low, Medium, and High settings. The remaining pathologies initially used MSW = 0.5 cm and were reoptimized with MSW = 1.0 cm. Plan quality, including total MU, delivery time, and dosimetric constraints, was statistically analyzed with a paired t-test. Results: Prostate plans exhibited the highest MU variation when changing “Fluence Smoothing” from Off to High (average ΔMU = −5.1%; P < 0.001). However, a High setting may increase overall MU when MSW = 0.5 cm. Gynecological plans changed substantially when MSW increased from 0.5 cm to 1.0 cm (average ΔMU = −29%; P < 0.001). Organs at risk sparing and planning target volumes remained within 1.2% differences. Replanning other pathologies with MSW = 1.0 cm affected breast and head and neck tumor plans (average ΔMU = −168.38, average Δt = −11.74 s, and average ΔMU = −256.56, average Δt = −15.05 s, respectively; all with P < 0.004). Radiosurgery and hypofractioned highly modulated plans did not yield statistically significant results. Conclusions: In breast, pelvis, head and neck, and prostate plans, starting with MSW = 1.0 cm optimally reduces MU and treatment time without compromising plan quality. MSW has a greater impact on MU than the “Fluence Smoothing” parameter. Plans with high modulation might present divergent behavior, requiring a case-specific analysis with MSW values higher than 0.5 cm.

Publisher

Medknow

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3