Overcoming Problems Caused by Offset Distance of Multiple Targets in Single-isocenter Volumetric Modulated Arc Therapy Planning for Stereotactic Radiosurgery

Author:

Ito Takaaki12,Kubo Kazuki1,Monzen Hajime1,Yanagi Yuya1,Nakamura Kenji1,Sakai Yusuke1,Nishimura Yasumasa3

Affiliation:

1. Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, Osakasayama, Osaka, Japan

2. Department of Radiological Technology, Kobe City Nishi Kobe Medical Center, Kobe, Hyogo, Japan

3. Department of Radiation Oncology, Faculty of Medicine, Kindai University, Osakasayama, Osaka, Japan

Abstract

Purpose: The purpose of the study is to investigate the impact of large target offset distances on the dose distribution and gamma passing rate (GPR) in single-isocenter multiple-target stereotactic radiosurgery (SIMT SRS) using volumetric modulated arc therapy (VMAT) with a flattening filter-free (FFF) beam from a linear accelerator. Methods: Two targets with a diameter of 1 cm were offset by ”±2, ±4, and ±6 cm from the isocenter in a verification phantom for head SRS (20 Gy/fr). The VMAT plans were created using collimator angles that ensured the two targets did not share a leaf pair from the multi-leaf collimator. To evaluate the low-dose spread intermediate dose spill (R50%), GPRs were measured with a criterion of 3%/2 mm using an electronic portal imaging device and evaluated using monitor unit (MU), modulation complexity score for VMAT (MCSv), and leaf travel (LT) parameters. Results: For offsets of 2, 4, and 6 cm, the respective parameters were: R50%, 4.75 ± 0.36, 5.13 ± 0.36, and 5.11 ± 0.33; GPR, 95.01%, 93.82%, and 90.67%; MU, 5893 ± 186, 5825 ± 286, and 5810 ± 396; MCSv, 0.24, 0.16, and 0.13; and LT, 189.21 ± 36.04, 327.69 ± 67.01, and 430.39 ± 114.34 mm. There was a spread in the low-dose region from offsets of ≥4 cm and the GPR negatively correlated with LT (r = −0.762). There was minimal correlation between GPR and MU or MCSv. Conclusions: In SIMT SRS VMAT plans with an FFF beam from a linear accelerator, target offsets of <4 cm from the isocenter can minimize the volume of the low-dose region receiving 10 Gy or more. During treatment planning, it is important to choose gantry, couch, and collimator angles that minimize LT and thereby improve the GPR.

Publisher

Medknow

Subject

Radiology, Nuclear Medicine and imaging,Biophysics

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3