Comparison of indirect cyanmethaemoglobin method and TrueHb haemometer point of care test with direct cyanmethaemoglobin (gold standard) method for strengthening anaemia detection at primary health care level

Author:

Khurana Rupinder1,Aggarwal Arun Kumar1,Thungapathra M.2,Sagar Vivek1

Affiliation:

1. Department of Community Medicine and School of Public Health, Postgraduate Institute of Medical Education and Research Chandigarh, India

2. Department of Biochemistry, Postgraduate Institute of Medical Education and Research Chandigarh, India

Abstract

ABSTRACT Introduction: The most reliable indicator for anaemia diagnosis at the population level is haemoglobin (Hb) estimation. The direct cyanmethaemoglobin method is considered the gold standard method for haemoglobin estimation. However, for resource constraint areas like primary health care (PHC) level, either blood samples are transported on filter paper for Hb testing (indirect cyanmethaemoglobin method) in laboratory or point of care testing is commonly used. Therefore, a comparative analysis of haemoglobin estimation of direct with indirect cyanmethaemoglobin method and also with TrueHb (Wrig Nanosystems Pvt. Ltd.) haemometer was done to strengthen anaemia diagnosis at the PHC level. Materials and Methods: This was a cross-sectional study. A total of 90 participants above 9 years of age, who visited the outpatient department (OPD) of health centre, Kheri and gave consent were included. Comparative analysis was done between Hb concentration assessed by indirect cyanmethaemoglobin method and TrueHb haemometer device against the gold standard method. Results: The mean Hb value estimated by direct method, TrueHb haemometer and indirect methods (filter paper A, B and C) was 11.42 ± 1.59 g/dl, 11.52 ± 1.54 g/dl, 10.66 ± 1.52 g/dl, 9.84 ± 1.50 g/dl and 10.19 ± 1.62 g/dl, respectively. There was no significant difference found between the mean Hb concentration estimated by the direct method and the TrueHb haemometer device. However, there was a significant difference in mean Hb values between the direct method and the indirect method. Therefore, regression analysis was done to estimate the correction factor for the indirect method. Conclusion: TrueHb metre device gave promising results in comparison to the gold standard method and can be used if resource permits in PHC centres. Indirect methods of haemoglobin estimation can be an alternative in resource-constraint settings, specifically for surveys. However, further studies are required for the validation of the indirect method.

Publisher

Medknow

Reference9 articles.

1. Global, regional, and national prevalence of anemia and its causes in 204 countries and territories, 1990–2019;Gardner;Current Developments in Nutrition. 2020 Jun;4(Supplement_2): 830

2. Symptomatology of anemia;Ludwig;Semin Oncol,2001

3. Precision, accuracy, and reliability of hemoglobin assessment with use of capillary blood;Morris;Am J Clin Nutr,1999

4. Recommendations for reference method for haemoglobinometry in human blood (ICSH standard 1995) and specifications for international haemiglobinocyanide standard (4th edition);Zwart;J Clin Pathol,1996

5. Comparison of haemoglobin estimates using direct and indirect cyanmethaemoglobin methods;Bansal;Indian J Med Res,2016

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3