Affiliation:
1. Advanced Eye Centre, PGIMER, Chandigarh, India
Abstract
Abstract
Purpose:
To perform an intraindividual comparison of the quality of vision and the effect of decentration between two aspheric intraocular lenses: aspheric balanced curve (ABC) design Vivinex iSert XY1 (Hoya Surgical Optics, Singapore) and anterior aspheric design Tecnis ZCB00 (Abbott Medical Optics, CA).
Setting:
Tertiary Eye Care Centre.
Design:
Prospective, randomized comparative study using a random number table.
Methods:
Thirty patients were randomized to the implantation of Vivinex iSert XY1 in one eye and Tecnis ZCB00 in the contralateral eye. Then, 12 weeks postoperatively, a laser ray-tracing aberrometer was used to evaluate the visual Strehl ratio, higher-order aberrations (HOA), decentration of IOL from the visual axis and geometric axis, angle alpha, and angle kappa. Contrast sensitivity was measured using the functional visual analyzer.
Results:
The visual internal Strehl ratio was higher (P < 0.05) at all pupil sizes and the spherical aberrations values were lower (P < 0.05) at larger pupil sizes (5 mm and 6 mm) in the Vivinex group. The mean decentration from the visual axis in the Vivinex group was significantly more than that in the Tecnis group (P < 0.01). With an increase in angle alpha, there was a greater decline in the visual Strehl ratio for Tecnis; however, there was a statistically significant decline at 3 mm pupil size for Tecnis ZCB00 (P = 0.028). The contrast sensitivity was similar for both IOLs.
Conclusion:
In comparison to an anterior aspheric design IOL, the ABC design IOL yielded better quality of vision, neutralized spherical aberrations to a larger extent, and provided a relatively superior quality of vision with decentration.