The purpose of this study is to examine the viability, precision, and consistency of a computer-based optokinetic nystagmus analyzer (nystagmus meter) for diagnosing eyesight in preschoolers. A total of 59 subjects who could pass the log of minimum angle of resolution (LogMAR) visual acuity chart were divided into three groups by age, 4-, 5-, and 6-year-old groups, and their visual acuity was tested with nystagmus meter. The percentage of children in each age group that could be detected by nystagmus was recorded along with the differences between these groups. The correlation between the test results from the two methods was found for each age group using the correlation coefficient method. Repeated measurements were used to assess the two visual acuity values of the measured nystagmus, and the repeatability of the two measurement techniques for different age groups was compared. The overall measurability of the visual acuity detected by nystagmus was 93.22%, and the measurability of the 4-, 5-, and 6-year-old groups was 90%, 95%, and 94.74%, respectively. There was no statistically significant difference in the measurability of subjects among all age groups (P = 1.0). The outcomes of the LogMAR visual acuity chart had a negative correlation with the visual acuity measured by the nystagmus meter. The overall correlation coefficient R value was -0.80, and the correlation coefficient R value of the 4-, 5-, and 6-year-old groups was -0.79, -0.76, and -0.87, respectively. The nystagmus meter has good feasibility, accuracy, and stability in visual acuity testing and can be used for visual acuity testing in children.

Publisher

Medknow

Subject

Ophthalmology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3