Comparing the Use of Smartphone and Vernier Calipers for Corneal Diameter Measurement in Nigerian Term Neonates

Author:

Mgbafulu N1,Ezisi C1,Ginger-Eke H1,Umeh R1

Affiliation:

1. Department of Ophthalmology, Alex Ekwueme Federal University Teaching Hospital, Abakaliki, Ebonyi State, Nigeria

Abstract

ABSTRACT Background: Corneal diameter (CD) measurement in newborns is essential for the detection of blinding ocular pathologies. It requires specially trained personnel, contact techniques, and the use of anesthesia. Smartphone use may obviate these challenges. Aim: This study compares CD measurement among newborns using a Smartphone and manual calipers. Subjects and Methods: This is a cross-sectional comparative study done at Alex Ekwueme Federal University Teaching Hospital, Abakaliki, Nigeria. Consecutively selected stable term neonates whose mothers consented to the study and met the inclusion criteria had their CDs measured with both calipers and smartphones and the values compared. Mean, range, and standard deviation were calculated using Statistical Package for the Social Sciences (SPSS) IBM (version 22), while paired t-test calculated the difference in mean CD measurements between both methods. Pearson’s correlation and Bland–Altman’s analysis determined the correlation and agreement between both methods, respectively. Inter and intraobserver repeatability were assessed. A P value less than 0.05 was taken as statistically significant. Results: The mean CD measurement using calliper method was 10.01 ± 0.29mm and 10.03 ± 0.24 mm right eye (RE) and left eyes (LE), respectively) for investigator 1 and 10.16 ± 0.25 mm and 10.19 ± 0.21 mm (RE, LE) for investigator 2. Using smartphone measurement, mean CD measurements of 9.98 ± 0.21 mm and 10.00 ± 0.29 mm (RE, LE) were observed for investigator 1 and 10.07 ± 0.19 mm and 10.06 ± 0.21 mm (RE, LE) for investigator 2. The difference in mean CD for both instruments and by both investigators was not statistically significant (P value > 0.05). Conclusion: Corneal diameter measurement in neonates using smartphone technology does not deviate significantly from caliper measurement; it is safe, easy to use, and advocated.

Publisher

Medknow

Subject

General Medicine

Reference23 articles.

1. Measurement of corneal diameter using smart phone technology;Harby;J Clin Exp Ophthalmol,2015

2. A review of corneal diameter, curvature and thickness values and influencing factors;Mashige;S Afr Optom,2013

3. Human corneal anatomy redefined:A novel pre-Descemet's layer (Dua's layer);Dua;Ophthalmology,2013

4. Recognising congenital glaucoma;Rudkin;Med J,2009

5. Microphthalmia, Anophthalmia and Coloboma and associated ocular and systemic features;Skalicky;JAMA Ophthalmol,2013