Comparative study on the quality of electrocardiogram and arrhythmia detection using wireless ambulatory Vigo SmartHeart Holter and conventional Holter

Author:

Bala Sudha1,Mirza Mehdi Ali2,Mavoori Archana3,Bandaru Rajiv Kumar4,Kampally Mallikharjuna5,Agarwal Sony5,Chormalle Rishabh4

Affiliation:

1. Department of Community Medicine, ESIC Medical College, Hyderabad, Telangana, India

2. Department of Pharmacology, ESIC Medical College, Hyderabad, Telangana, India

3. Department of Dermatology, ESIC Medical College, Hyderabad, Telangana, India

4. Department of General Medicine, ESIC Medical College, Hyderabad, Telangana, India

5. Medical Affairs, Vigocare, Hyderabad, Telangana, India

Abstract

Abstract Background: Wireless Holter monitors can be a great alternative to the traditional wired holter monitoring potentially giving better patient outcomes. However, the technology is new and less explored for the quality of ECG wave forms. Objective: This study was taken up to compare the quality of electrocardiogram (ECG) and arrhythmia detection using a traditional Holter and a wireless ambulatory smart ECG – Vigo SmartHeart (VSH) that functions with a Mobile application. Materials and Methods: The present study was cross-sectional in nature. We enrolled the patients indicated for continuous ECG monitoring. All the enrolled patients were connected to VSH wireless Holter solution and a conventional Holter monitor simultaneously. The ECGs generated were compared through descriptive statistics such as mean, standard deviation, and correlation by Pearson’s correlation method. Results: Seventy-one patients had simultaneous ECG recordings by both systems. The Pearson correlation for P wave, QRS, PR, QT, and ST was measured as 0.93, 0.93, 0.99, 0.36, and 0.98, respectively. The minimum heart rate (HR), maximum HR, RR interval, P wave, QRS complex, PR, and ST segments showed a statistically significant correlation between Smart Heart and Conventional Holter. 17 ECG events were reported, of which 14 were detected by both. The events missed by conventional Holter and detected by VSH were Atrioventricular block (2), Supraventricular tachycardia (1), and Intraventricular conduction delay (1). Conclusions: The quality of the report generated by both the methods was comparable to ECG wave quality, However, arrhythmia was picked up by the SmartHeart with better precision compared to conventional Holter. Thus, VSH can serve as a viable and potential alternative tool to screen and help in reducing the morbidity and mortality related to arrhythmias.

Publisher

Medknow

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3