Off-label use of KONAR-MF™ occluder for transcatheter closure of patent ductus arteriosus in unusual circumstances

Author:

Banpurkar Ashishkumar1,Koneti Nageswara Rao2,Thakur Prashant3,Kulkarni Snehal M.1,Prabhu Atul3,Doniparti Pavan Kumar Venkata2

Affiliation:

1. Department of Pediatric Cardiology, Sri Sathya Sai Sanjeevani Centre for Child Heart Care, Navi Mumbai, Maharashtra, India

2. Department of Pediatric Cardiology, Rainbow Children’s Heart Institute, Hyderabad, Telangana, India

3. Department of Pediatric Cardiology, Sri Sathya Sai Sanjeevani Center for Child Heart Care, Chhattisgarh, India

Abstract

ABSTRACT Background: Transcatheter patent ductus arteriosus (PDA) closure with unusual morphology associated with multiple constrictions or unusual circumstances such as interruption of inferior vena cava (IVC) is still challenging. This study evaluates the use of KONAR-MF™ (Lifetech Scientific Co Ltd., Shenzhen) for transcatheter closure of PDA in such situations. Materials and Methods and Results: Between January 2021 and October 2021, 24 patients from three different tertiary care centers underwent PDA device closure using the KONAR-MF™ device and are included in the study. Patient demographics, echocardiographic assessment, procedural details including the approach, device details, complications, and follow-up data were recorded. The median age and weight of the cohort were 2 years (0.5–41) and 10 kg (5–98), respectively. The indications for using KONAR-MF™ were (a) unusual morphology of PDA in 14, long tubular PDA in 8, and long tubular with multiple constrictions in 6 patients and (b) unusual circumstances such as interrupted IVC in 6 and inability to cross PDA antegradely in 4 due to dilated main pulmonary artery. The procedure was successful in all the patients and no procedure-related complications were noted in this cohort. Median follow-up of 9 months (range: 5–18) showed no residual shunt and only one patient had flow acceleration across the left pulmonary artery (peak Doppler gradient 16 mmHg). Conclusions: KONAR-MF™ device can be used safely and effectively for transcatheter closure of PDA with unusual anatomy or in unusual circumstances.

Publisher

Medknow

Subject

Cardiology and Cardiovascular Medicine,Pediatrics, Perinatology and Child Health

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司