Karthhik Raadhikka,Raj Bharath,Surendra Kumar G. P.
Abstract
Statement of Problem:
Numerous different attachments are used to retain overdentures. Hence, there is a need to assess new chairside attachment system based on polyvinylsiloxane (Retention.sil [R.S]) to know its in vitro behavior to be used clinically.
Purpose:
To assess and compare the retentive force of two female matrix retention system (R.S with a conventional O-ring [OR] attachment system) in tooth-supported overdenture.
Materials and Methods:
Completely edentulous mandibular model with two canine root-anchored ball attachments for overdentures were fabricated. A total of 30 mandibular dentures were made and divided to form two groups, each with a different attachment design. One group (test) received 5 pairs of R.S attachments while the other group (control) received 15 pairs of ORs. Retention force was measured by subjecting them to 540 cycles of insertion and de-insertion cycles (representing 6 months functional life) in axial direction before and after thermocycling.
Results:
Statistical analysis comprised Kolmogorov–Smirnov test, Friedman test, and Wilcoxon signed-rank test (α = 0.05). The initial retention of R.S attachments was 5.90 ± 0.77 N and of ball/OR 17.36 ± 3.81 N. The final retention of R.S attachments was 5.41 ± 0.72 N and of ball/OR 12.91 ± 3.18 N. The OR developed higher retentive force as compared to the R.S attachments (P < 0.05). However, no significant changes in retention force were observed for R.S groups after repeated dislodging and thermal undulation.
Conclusion:
Both systems presented acceptable retention capacities after 540 cycles. The newly developed R.S attachments can provide an alternative to OR attachments for short-term retention. Clinical studies are however required to elucidate the long-term performance of these materials.
Reference35 articles.
1. Comparison of retention and strain energies of stud attachments for implant overdentures;Petropoulos;J Prosthodont,2011
2. A comparative in vitro study on the retention and stability of implant-supported overdentures;Sadig;Quintessence Int,2009
3. Effects of clip materials on stress distribution to maxillary implant overdentures with bar attachments;Tanoue;J Prosthet Dent,2016
4. Soft liner-retained, implant-supported overdenture: A technical note;Cain;Int J Oral Maxillofac Implants,1998
5. Effect of abutment angulation in the retention and durability of three overdenture attachment systems: An in vitro study;Aroso;J Adv Prosthodont,2016
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献