Blind spot in ultrasound central corneal thickness measurement – Central corneal thickness of apex versus central corneal thickness of vertex

Author:

Ramesh Prasanna Venkatesh1,Parthasarathi Sathyan2,Ashok Abhinay3,John Rajesh Kumar4

Affiliation:

1. Medical Officer, Department of Glaucoma and Research, Mahathma Eye Hospital Private Limited, Tiruchirappalli, Tamil Nadu, India

2. Director, Sathyan Eye Care Hospital and Coimbatore Glaucoma Foundation, Coimbatore, Tamil Nadu, India

3. Consultant, Abhi Nethra Eye Care, Udupi, Karnataka, India

4. Biostatistician, Sathyan Eye Care Hospital and Coimbatore Glaucoma Foundation, Coimbatore, Tamil Nadu, India

Abstract

Aims: This study aimed to compare central corneal thickness (CCT) using ultrasound pachymetry (USP) (CCT apex) and specular microscopy (CCT vertex) and also to find out the intra-reading variability from the two instruments. Settings and Design: A prospective, observational study was conducted in a tertiary eye care center in southern India. Materials and Methods: This study was conducted on 12 patients (24 eyes, 96 data set eyes, 768 data points in total) aged 20–50, in non-pathological corneas. Eight CCT measurements by specular microscopy and eight measurements by USP were taken by two different experienced observers on day 1, followed by repeating the same on day 2. Statistical Analysis: The readings were averaged and compared by paired t-test. SPSS software was used for statistical analysis. Variability between the eight readings produced by the same instrument was calculated and coefficient of variation was plotted. Inter-examiner variability and intra-examiner variability of the two modalities were studied, and it was considered statistically significant if P < 0.05. Results: The mean CCT by USP was 522.71 µm and the mean CCT by specular microscopy was 519.43 µm. The coefficient of variation of the eight readings varied significantly between the machines. On an average, the coefficient of variation was 0.4% compared to 0.8% in specular microscopy and USP, respectively. Conclusion: The intra-reading variability of ultrasound CCT is twice as that compared to specular microscopy. The clinical relevance of this spread is relevant and important in modern-day practice. Hence, it is better advisable that CCT readings from both the modalities, not be used interchangeably.

Publisher

Medknow

Subject

General Earth and Planetary Sciences,General Environmental Science

Reference28 articles.

1. Measurement of central corneal thickness using ultrasound pachymetry and Orbscan II in normal eyes;Sadoughi;J Ophthalmic Vis Res,2015

2. Measurement of central corneal thickness by ultrasonic pachymeter and oculus pentacam in patients with well-controlled glaucoma: Hospital-based comparative study;Rashid;Clin Ophthalmol,2016

3. Repeatability and agreement of CCT measurement in myopia using entacam and ultrasound pachymetry;Luo;Int J Ophthalmol,2012

4. Corneal thickness measurements with the Topcon SP-2000P Specular Microscope and an ultrasound pachymeter;Bovelle;Arch Ophthalmol,1999

5. Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, Orbscan and ultrasound pachymetry;Wong;Eye (Lond),2002

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3