Comparative evaluation of self-pressurized Air-Q® and Proseal™ LMA® in patients undergoing elective surgery under general anaesthesia: A randomized clinical trial

Author:

Rana Savita1,Anand Lakesh Kumar2,Singh Manpreet1,Kapoor Dheeraj1,Gupta Deepika1,Kaur Harpreet1

Affiliation:

1. Department of Anaesthesia and Intensive Care, Government Medical College and Hospital, Chandigarh, India

2. Department of Anaesthesia, Maharishi Markandeshwar Institute of Medical Sciences and Research, Mullana, Ambala, Haryana, India

Abstract

Abstract Background and Aims: Intra-cuff pressure of Air-Q self-pressurized laryngeal airways (Air-Q SP) balances airway pressure and adapts to patient’s pharyngeal and periglottic structures, thus improves oropharyngeal leak pressure (OLP).This study was performed to compare efficacy of Air-Q SP with Proseal laryngeal mask airway (PLMA) in patients undergoing elective surgery. Material and Methods: The study design was prospective, randomized and controlled. Ninety patients were randomly assigned to Air-Q SP or PLMA group. All patients were premedicated and shifted to operation theatre. Monitoring was instituted. After securing IV-line, induction with inj. Morphine + Propofol, relaxation with inj. Vecuronium was done. Supraglottic was inserted according to group allocation. Outcome measures were OLP, fibreoptic view of larynx, success rate, device insertion parameters, haemodynamic and respiratory parameters and post-operative laryngopharyngeal complications. Neostigmine + glycopyrrolate were given, device was extubated. Results: All supraglottic airway devices (SADs) were successfully placed in two attempts. The mean initial OLP, OLP at 10 minutes, and device insertion time were significantly lower in Air-Q SP group. Fiber-optic laryngeal view grading was significantly better with Air-Q SP. No significant difference was observed with respect to rate of successful insertion in first attempt, ease of insertion, and manipulations required. The hemodynamic/respiratory parameters and post-operative sore throat in the two both groups were similar. Conclusions: Proseal LMA has a higher OLP than Air-Q SP but average insertion time was better, and fiber-optic grading of laryngeal view was shorter with Air-Q SP. However, Air-Q SP and Proseal LMA were both effective for lung ventilation.

Publisher

Medknow

Reference21 articles.

1. Prospective evaluation of the self-pressurized air-Q intubating laryngeal airway in children;Jagannathan;Paediatr Anaesth,2011

2. The air-Q intubating laryngeal airway vs the LMA:Proseal™:A prospective randomized trial of airway seal pressure;Galgon;Anaesthesia,2011

3. Comparison of the performance of the self-pressurized air-Q intubating laryngeal airway with the LMA-ProSeal in pediatric patients under general anesthesia:A randomized controlled trial;Aly;Ain Shams J Anaesthesiol,2019

4. A clinical evaluation of the intubating laryngeal airway as a conduit for tracheal intubation in children;Jagannathan;Anesth Analg,2011

5. A randomized crossover comparison between the Laryngeal Mask Airway-Unique™ and the air-Q intubating laryngeal airway in children;Jagannathan;Paediat Anaesth,2012

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献