Abstract

Introduction: Small animal models are widely used in basic research. However, experimental systems requiring extracorporeal circuits are frequently confronted with limitations related to equipment size. This is particularly true for oxygenators in systems with limited volumes. Thus we aimed to develop and validate an ultra mini-oxygenator for low-volume, buffer-perfused systems. Methods: We have manufactured a series of ultra mini-oxygenators with approximately 175 aligned, microporous, polypropylene hollow fibers contained inside a shell, which is sealed at each of the two extremities to isolate perfusate and gas compartments. With this construction, gas passes through hollow fibers, while perfusate circulates around fibers. Performance of ultra mini-oxygenators (oxygen partial pressure (PO2), gas and perfusate flow, perfusate pressure and temperature drop) were assessed with modified Krebs-Henseleit buffer in an in vitro perfusion circuit and an ex vivo rat heart preparation. Results: Mean priming volume of ultra mini-oxygenators was 1.2±0.5 mL and, on average, 86±6% of fibers were open (n=17). In vitro, effective oxygenation (PO2=400–500 mmHg) was achieved for all flow rates up to 50 mL/min and remained stable for at least 2 hours (n=5). Oxygenation was also effective and stable (PO2=456±40 mmHg) in the isolated heart preparation for at least 60 minutes (“venous” PO2=151±11 mmHg; n=5). Conclusions: We have established a reproducible procedure for fabrication of ultra mini-oxygenators, which provide reliable and stable oxygenation for at least 60–120 min. These oxygenators are especially attractive for pre-clinical protocols using small, rather than large, animals.

Publisher

SAGE Publications

Subject

Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3