A Survey of Membrane Oxygenator Heat - Exchanger Integrity Testing at Cardiac Surgery Centres in Great Britain and Ireland

Author:

Carlton Matthew1,Campbell John1

Affiliation:

1. Department of Clinical Perfusion Science, Nottingham University Hospitals Trust, Nottingham - UK

Abstract

Membrane oxygenator heat exchanger (HE) device failure is reported to be very low for both short-and long-term extracorporeal devices. All oxygenator manufacturers provide instructions for leak testing of their HE devices prior to patient use. In addition to these recommendations, since 2006 at Nottingham University Hospitals (NUH) we have also additionally pressure tested HE devices prior to use. We conducted a national survey of cardiac centers in Great Britain and Ireland to determine the methods undertaken in individual centers for validation of the integrity of HE devices. Furthermore, we also collected information on the routine maintenance techniques utilized within these centers to inhibit microbial growth in the water used in the heater-cooler units (HCUs). In total, 34 responses were collected from the 57 centers performing cardiac surgery, producing a response rate of 60%. Of the responding centers, 71% are adhering to manufacturer's recommended guidelines of circulating the water through the device for 5 minutes. Of these centers, 17% reported detecting a leak between the HE and membrane compartment of the oxygenator. In responding centers, 29% reported using the pressure test technique. In the centers utilizing pressure testing, 60% reported detecting a leak. This survey reports an association of a greater HE leak detection rate using the pressure test technique compared to using water testing in isolation (p = 0.034). We believe the pressure testing method provides the perfusionist with confidence in the integrity of the HE for short- and long-term circulatory support devices prior to use in both elective and emergency situations.

Publisher

SAGE Publications

Subject

Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3