IABP Timing and Ventricular Performance - Comparison between a Compliant and a Stiffer Aorta: A Hybrid Model Study Including Baroreflex

Author:

Fresiello Libera12,Khir Ashraf W.3,Di Molfetta Arianna1,Kozarski Maciej2,Ferrari Gianfranco1

Affiliation:

1. Section of Rome, Institute of Clinical Physiology, National Research Council, Rome - Italy

2. Nałętecz Institute of Biocybernetics and Biomedical Engineering, Polish Academy of Sciences, Warsaw - Poland

3. Brunel Institute for Bioengineering, Brunel University, Uxbridge, Middlesex - UK

Abstract

Introduction The aim of this study was to investigate the effects of the intra aortic balloon pump (IABP) and of aortic compliance on left ventricular performance, including the effects of baroreflex control. Methods The study was conducted using a hybrid cardiovascular simulator, including a computational cardiovascular sub-model, a hydraulic sub-model of the descending aorta, and a baroreflex computational sub-model. A 40 cc balloon was inserted into a rubber tube component of the hydraulic sub-model. A comparative analysis was conducted for two aortic compliances (C1 = 2.4 and C2 = 1.43 cm3/mmHg, corresponding to an aortic pulse pressure of 23 mmHg and 35 mmHg, respectively), driving the balloon for different trigger timings. Results Under C1 conditions, the IABP induced higher effects on baroreflex activity (decrement of sympathetic efferent activity: 10% for C1 and 14.7% for C2) and ventricular performance (increment of cardiac output (CO): 3.7% for C1 and 5.2% for C2, increment of endocardial viability ratio (EVR): 24.8% for C1 and 55% for C2). The best balloon timing was different for C1 and C2: inflation trigger timing (from the dicrotic notch) -0.09 s for C1 and -0.04 s for C2, inflation duration 0.25 s for C1 and 0.2 s for C2. Conclusions Early inflation ensures better EVR, CO, and an increment of the afferent nerve activity, hence causing peripheral resistance and heart rate to decrease. The best balloon timing depends on aortic compliance, thus suggesting the need for a therapy tailored to the specific conditions of individual patients.

Publisher

SAGE Publications

Subject

Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3