Affiliation:
1. Clinical Orthopedics and Traumatology Department, Catholic
University of the Sacred Heart, Rome - Italy
Abstract
Background Metallic ball heads for total hip replacement (THR) bearings with ceramicized surfaces were introduced in orthopedics during the second half of the 1980s, with the aim of decreasing the wear of polyethylene cups. Methods An analysis was made of the literature regarding outcomes for metallic ball heads with ceramicized surfaces now in clinical use (TiN, TiNbN, ZrN, monoclinic ZrO2), as well as carbon coatings (pyrolytic carbon, diamond-like carbon) and silicon nitride as coatings in ball heads for THR bearings. Results Notwithstanding the diffusion of ceramicized ball heads in THRs, there are few reports about their clinical outcomes in hip arthroplasty. In addition, several clinical studies and some registry data are putting under scrutiny the clinical advantages of ceramicized ball heads over cobalt chrome (CoCr) alloy and ceramic ball heads. Conclusions The wear of THR bearings with ceramicized ball heads looks like it depends more on the behavior of the polyethylene cups than on the treatment of the ball head surface. The risk of coating damage and of its consequences has to be taken into account in selecting this type of bearing.
Subject
Biomedical Engineering,Biomaterials,General Medicine,Bioengineering,Biophysics
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献