The Value of Serial Metal Ion Levels in following up Patients with Metal-on-Metal Hip Arthroplasty

Author:

McHugh Gavin1,Merchant Rajiv1,Kelly Gabrielle E.2,Bergin Karen M.1,McCoy Gerald F.1,Wozniak Andrew P.1,Quinlan John F.1

Affiliation:

1. Lourdes Orthopaedic Hospital, Kilcreene, Kilkenny - Ireland

2. School of Mathematical Sciences, University College Dublin, Dublin - Ireland

Abstract

Background The sensitivity of cobalt (Co) and chromium (Cr) ion-levels in detecting poorly performing metal-on-metal hip implants is low. This study proposes that serial changes in ion-levels are a more accurate marker of arthroplasties at risk. Methods Serial metal ion-levels and implant data of 285 patients with ASR resurfacing or replacement were studied. Patient and implant characteristics were analysed using univariate and multivariate analyses. Results 111 (39%) had revision surgery. Time since index surgery (p≤0.001), acetabular inclination (p≤0.001), their interaction (p≤0.001) and femoral head size (p = 0.01) were significant variables. Head size (≤51 mm) had lower Co and Cr levels (p = 0.01). Cr/Co showed marginal decrease over time in the unrevised group and no decrease prerevision. Conclusions Repeated measurement of ion-levels were higher in the revision group suggesting that serial measurements rather than absolute values may have a role to play in predicting implant failure.

Publisher

SAGE Publications

Subject

Orthopedics and Sports Medicine,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3