Viabahn stent for hemodialysis shunt: efficacy, long segment recanalization and prognostic factors for reintervention

Author:

Salsano Giancarlo1,Trezzi Matteo2,Barattini Matteo1,Puccianti Franca1,Romano Nicola1,Zattera Tito2,Chiappini Nadia2,Londrino Francesco3,Rolla Davide2,Stefanini Teseo1

Affiliation:

1. Unit of Radiology and Interventional Radiology, General Hospital Sant’Andrea, La Spezia - Italy

2. Departments of Nephrology and Dialysis, General Hospital Sant’Andrea, La Spezia - Italy

3. Departments of Nephrology and Dialysis, General Hospital Sant’Eugenio, Rome - Italy

Abstract

Introduction: The study evaluated the Heparin Bioactive Surface (HBS) Viabahn Stent (W.L. Gore & Associates, Flagstaff, Arizona) efficacy in the maintenance or re-establishment of hemodialysis. Materials and methods: Fifty HBS Viabahn stents deployed in 37 consecutive patients with hemodialysis dysfunction from January 2008 to May 2016 were evaluated in a single-institution retrospective review. Outcomes were stent patency intended as primary circuit patency (PP), assisted primary patency (APP), target lesion primary patency (TLPP) and secondary patency (SP). Moreover, the risk factor analysis for hemodialysis dysfunction that required reintervention was performed. A subgroup analysis was conducted to assess patency of Viabahn stent to treat peripheral venous long segment obstruction (LSO). Results: Overall Kaplan-Meyer PPs were 60% at 12 months and 42% at 24 months. Overall TLPP estimated rates were 68% and 49% at 12 and 24 months, respectively. The corresponding SP rates were 85% and 78% at the same period. Estimated PP rates at 12 and 24 months for stent placement after peripheral venous long segment recanalization procedure were 53% and 31%, respectively. Corresponding SP rates were 82% and 68%, respectively. The APP rates were 79% at 12 months and 61% at 24 months. Female sex, access age and thrombosis were associated with reduced primary patency. Conclusions: Considering the high rates of PP, TLPP, APP and SP, Viabahn stents have been proven effective in maintaining or re-establishing the hemodialysis access. Moreover, stent placement after recanalization of LSO of venous out-flow represented a valid approach to rescue a dysfunctional fistula that would otherwise be abandoned.

Publisher

SAGE Publications

Subject

Nephrology,Surgery

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3