Abstract
The Harrison–Nicolle Intramedullary Peg is used for the arthrodesis of digital joints. We report our experience of complications found using this peg in a series of 16 arthrodeses. At review after an average of 2.9 years, three cases had been revised. Only five of the remaining 13 cases showed a bony fusion and in only two of these five was the fusion in an acceptable position. In nine cases the joint became straight or extended; the preset angle of the peg did not determine the final angle of the arthrodesis. In five cases the tips of the peg appeared to be eroding through bone.