Affiliation:
1. From the Department of Otolaryngology-Head and Neck Surgery, University of California, Irvine, California
2. School of Engineering, Harvey Mudd College, Claremont, CA
Abstract
OBJECTIVE: To evaluate and optimize the design of a removable and inexpensive internal stabilization device to reduce the effect of intention tremor during laryngeal microsurgery. STUDY DESIGN AND SETTING: In this laboratory investigation, stabilizers were designed and constructed to allow a nonobstructing view of the surgical field, permit simple insertion and removal, and accommodate microsurgical instruments. Prototype stabilizers were tested by using a Dedo laryngoscope, a measurement grid, and video recording equipment, which recorded instrument tremor within the magnified operative field for later analysis. Physicians also rated instrument stability, mobility, visualization, and ease of use on a survey form. RESULTS: Instrument tremor was reduced approximately 90%, with little obstruction of view of the surgical field. Instrument range of motion was reduced but improved rapidly as the stabilizer bar was moved further from the tip of the laryngoscope. CONCLUSIONS: Use of a stabilization device in the laryngoscope lumen reduces instrument tremor and has the potential to improve surgical performance during laryngeal microsurgery. EBM rating: B-3.
Subject
Otorhinolaryngology,Surgery
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献