Author:
G Rajeshkumar,H Mohammed Nasrullah,S Nithesh Kanna,Kumar Santhosh,M Vignesh
Abstract
In a hip replacement procedure, the damaged bone and cartilage are removed and replaced with an artificial component known as prosthesis. Despite advancements in implant sterilization, design, fixation techniques, and the introduction of robotic surgery, a persistent challenge is to identify an optimal, patient-specific hip implant that meets individual criteria. The primary objective of the proposed study was to create a highly accurate patient-specific hip implant by standardizing the existing design. The secondary objective aimed to demonstrate the superiority of a customized design over a conventional one. Geometric measurements of the hip were extracted from CT scans using MIMICS 20.0 software, and the implant design was developed using SolidWorks. Finite Element Analysis (FEA) was employed for meshing and analyzing the planned implant. Comparative research through FEA analysis indicated that a customized implant made from SSL 13 material outperformed the standard implant, showcasing its suitability for the patients studied.