Agreement between i-STAT and YSI 2300 devices to determine lactate concentrations in dogs undergoing exercise

Author:

Berkman C.1,Pereira M.C.2,Nardi K.B.1,Pereira G.T.3,Soares O.A.B.4,Restan W.A.Z.1,Queiroz-Neto A.1,Ferraz G.C.1

Affiliation:

1. Laboratory of Pharmacology and Equine Exercise Physiology (LAFEQ), College of Agriculture and Veterinary Sciences (FCAV), Department of Animal Morphology and Physiology, UNESP Univ Estadual Paulista, 14884-900 Jaboticabal, São Paulo, Brazil.

2. Universidade Federal do Mato Grosso – Campus Sinop, 78555000 Sinop, Brazil.

3. Department of Exact Sciences, UNESP Univ Estadual Paulista, 14884-900 Jaboticabal, São Paulo, Brazil.

4. Brazilian Army, 2° Batalhão de Polícia do Exército, 06236100 São Paulo, Brazil.

Abstract

Little information is available comparing the i-STAT and the YSI 2300 Stat Plus devices to determine the lactate concentration [Lac] in dogs undergoing intense exercise. The reproducibility of the YSI 2300 for quantifying the [Lac] in canine blood [Lac]b and plasma [Lac]p samples has been observed. In addition, the i-STAT handheld device was used to quantify [Lac] in dogs subjected to exercise, and the results were compared with that of YSI 2300. Venous blood samples of Beagle and American Pit Bull Terrier dogs were obtained during an intense exercise training on a treadmill. [Lac]p and [Lac]b were quantified using the YSI 2300 instrument to determine the reproducibility of the results. A total of 52 specimens were compared for both plasma and whole blood. For comparing the devices (YSI 2300 vs i-STAT), 96 samples were used. Ordinary least products regression, the correlation coefficient, and Bland-Altman plots were used to assess the agreement of using the i-STAT device. The relationship between duplicate measurements of both [Lac]b and [Lac]p by YSI 2300 was strong (r=0.99). A correlation between the data obtained using the i-STAT and YSI 2300 instruments was observed for both the [Lac]p (r=0.97) and [Lac]b (r=0.88). The i-STAT exhibited a small constant bias (-0.25 mmol/l) compared to YSI 2300 ([Lac]b). There were proportional biases of 0.89 mmol/l for [Lac]p and 1.22 mmol/l for [Lac]b when using YSI 2300 vs i-STAT. We confirmed the reproducibility of the YSI 2300 for canine lactate blood/plasma samples. The results obtained by the i-STAT and YSI 2300 analyser were highly correlated, but a small constant bias was observed between them. The i-STAT device can be used in clinical evaluations, and it is also adequate for designing and monitoring fitness programmes.

Publisher

Wageningen Academic Publishers

Subject

Physiology (medical),Veterinary (miscellaneous),Orthopedics and Sports Medicine,Physiology,Biochemistry,Endocrinology, Diabetes and Metabolism,Biophysics

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3