Affiliation:
1. Department of Chemistry and Biochemistry, Northern Illinois University, DeKalb, Illinois 60115
Abstract
A quartz acousto-optic tunable filter (AOTF) is utilized as a wavelength selection device for inductively coupled plasma atomic emission spectrometry (ICP-AES). System design is detailed. Fundamental characterizations indicate that the useful spectral range exceeds the 206 to 405 nm range. Wavelength calibration is extremely accurate with the correlation coefficient of the ICP atomic emission wavelength/inverse frequency plot exceeding 0.999999. Bandpasses were measured as 0.06 nm at 226 nm and 0.36 nm at 405 nm. These values are comparable to those of a moderate-resolution monochromator. The system was characterized with 12 metals: Ag, Ba, Ca, Cd, Cr, Cu, Eu, Fe, Hg, Mg, Mn, and Pb. Detection limits are in the single to tens of ng/mL range for many metals. Linearity is demonstrated from lower concentration ranges near the detection limit.
Subject
Spectroscopy,Instrumentation
Reference17 articles.
1. Goutzoulis A. P., Pape D. R., Design and Fabrication of Acousto-Optic Devices (Marcel Dekker, New York, 1994), pp. 197–285.
2. Performance and Selected Applications of an Acousto-Optic Tunable Filter Near-Infrared Spectrometer
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献