Abstract
The transfer of a calibration model for determining fiber content in flax stem was accomplished between two near-infrared spectrometers, which are the same brand but which require a standardization. In this paper, three factors, including transfer sample set, spectral type, and standardization method, were investigated to obtain the best standardization result. Twelve standardization files were produced from two sets of the transfer sample (sealed reference standards and a subset of the prediction set), two types of the transfer sample spectra (raw and preprocessed spectra), and three standardization methods (direct standardization (DS), piecewise direct standardization (PDS), and double window piecewise direct standardization (DWPDS)). The efficacy of the model transfer was evaluated based on the root mean square error of prediction, calculated using the independent prediction samples. Results indicated that the standardization using the sealed reference standards was unacceptable, but the standardization using the prediction subset was adequate. The use of the preprocessed spectra of the transfer samples led to the calibration transfers that were successful, especially for the PDS and the DWPDS correction. Finally, standardization using the prediction subset and their preprocessed spectra with DWPDS correction proved to be the best method for transferring the model.