Noninvasive Near-Infrared Blood Glucose Monitoring Using a Calibration Model Built by a Numerical Simulation Method: Trial Application to Patients in an Intensive Care Unit

Author:

Maruo Katsuhiko1,Oota Tomohiro1,Tsurugi Mitsuhiro1,Nakagawa Takehiro1,Arimoto Hidenobu1,Hayakawa Mineji1,Tamura Mamoru1,Ozaki Yukihiro1,Yamada Yukio1

Affiliation:

1. Matsushita Electric Works Ltd., Kadoma, Osaka 571-8686, Japan (K.M., T.O., M.Tsurugi, T.N.); National Institute of Advanced Industrial Science and Technology, Tsukuba 305-8564, Japan (H.A.); ICU, Hokkaido University Hospital, Sapporo 060-8648, Japan (M.H.); Division of Cellular Life Science, Hokkaido University, Sapporo 060-0812, Japan (M.Tamura); Department of Chemistry and Research Center for Near-Infrared Spectroscopy, School of Science and Technology, Kwansei-Gakuin University, Sanda 669-1337, Japan ...

Abstract

We have applied a new methodology for noninvasive continuous blood glucose monitoring, proposed in our previous paper, to patients in ICU (intensive care unit), where strict controls of blood glucose levels are required. The new methodology can build calibration models essentially from numerical simulation, while the conventional methodology requires pre-experiments such as sugar tolerance tests, which are impossible to perform on ICU patients in most cases. The in vivo experiments in this study consisted of two stages, the first stage conducted on healthy subjects as preliminary experiments, and the second stage on ICU patients. The prediction performance of the first stage was obtained as a correlation coefficient (r) of 0.71 and standard error of prediction (SEP) of 28.7 mg/dL. Of the 323 total data, 71.5% were in the A zone, 28.5% were in the B zone, and none were in the C, D, and E zones for the Clarke error-grid analysis. The prediction performance of the second stage was obtained as an r of 0.97 and SEP of 27.2 mg/dL. Of the 304 total data, 80.3% were in the A zone, 19.7% were in the B zone, and none were in the C, D, and E zones. These prediction results suggest that the new methodology has the potential to realize a noninvasive blood glucose monitoring system using near-infrared spectroscopy (NIRS) in ICUs. Although the total performance of the present monitoring system has not yet reached a satisfactory level as a stand-alone system, it can be developed as a complementary system to the conventional one used in ICUs for routine blood glucose management, which checks the blood glucose levels of patients every few hours.

Publisher

SAGE Publications

Subject

Spectroscopy,Instrumentation

Cited by 36 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Non-Destructive Sensor for Glucose Solution Concentration Detection Using Electromagnetic Technology;Micromachines;2024-06-05

2. Non-invasive blood glucose measurement based on visible light and embedded system;2021 International Conference on Optical Instruments and Technology: Optoelectronic Measurement Technology and Systems;2022-07-08

3. Recent advances in optical sensors for continuous glucose monitoring;Sensors & Diagnostics;2022

4. On the use of fiber lasers in non-invasive blood glucose monitoring;Optical Fiber Technology;2022-01

5. Accuracy of Blood Glucose Measurement and Blood Glucose Targets;Journal of Diabetes Science and Technology;2020-02-11