Abstract
Modifications affecting the resolution and frequency calibration capability of the Beckman IR-9 infrared spectrophotometer are described and discussed. Under favorable conditions these modifications enable the instrument to be calibrated with an absolute accuracy of 0.03 cm−1 or better; resolution is also improved so that spectral slitwidths may be reduced by approximately √2 over normal specifications. These results are illustrated with vibration-rotation spectra from NH3, CH4, and HBr.
Subject
Spectroscopy,Instrumentation
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. The infrared spectra of HCF3 and DCF3;Journal of Molecular Spectroscopy;1975-10
2. The infrared spectrum of HC13F3;Journal of Molecular Spectroscopy;1975-10