Affiliation:
1. Engineering Research Center, Western Electric Company, Incorporated, Princeton, New Jersey 08540
Abstract
Tantalum–aluminum thin film composition has been determined destructively by atomic absorption and nondestructively by x-ray fluorescence spectroscopy. Samples representing several compositions (20–80 at.% aluminum) and thicknesses (500–6000 Å) were sputtered on glass, graphite, and platinum substrates. The films were dissolved from the platinum substrates for the determination of aluminum by atomic absorption. The weights of tantalum per unit area obtained by difference using this destructive technique were applied to the same samples on glass substrates for correlation with nondestructive x-ray fluorescence measurements. A linear curve, which is free from enhancement and absorption effects, is obtained for tantalum. This curve relates the nondestructive fluorescence intensities to film weights (µg/cm2) of tantalum. The composition of the film is determined nondestructively by utilizing this curve and the total film weight which is obtained by weighing the substrate before and after sputtering. Alternately, composition may be determined destructively by atomic absorption utilizing films dissolved from platinum substrates
Subject
Spectroscopy,Instrumentation
Reference9 articles.
1. Gasometric Determination of Nitrogen in Refractory Metal Nitrides of Groups IVB, VB, and VIB.
2. Handbook of Chemistry and Physics, edited by Hodgman C. D. (Chemical Rubber Publ., Cleveland, 1960, 1961), 42nd ed., pp. 666–667.
3. Treatise on Analytical Chemistry, 6, edited by Kolthoff I. M., Elving P. J., and Sandell E. B. (Wiley, New York, 1964) Pt. 2, p. 222.
4. Physical Measurement and Analysis of Thin Films