Affiliation:
1. Hosler Research Center, Applied Research Laboratories, Goleta, California 93017
Abstract
A direct current capillary arc operating in argon at atmospheric pressure has been combined with a device which produces fine particles directly from the surface of a solid metal sample. The aerosol generator is a simple device in which the sample acts as the cathode of a dc arc discharge. Rapid movement of the cathode spot produces uniform sampling over a well-defined area. A flowing gas stream transports the aerosol particles to the capillary arc discharge. Reproducibility and analytical curves are shown for ten elements in low alloy, stainless, and tool steel samples, for total analytical times of 10 sec.
Subject
Spectroscopy,Instrumentation
Reference5 articles.
1. Direct Nebulization of Metal Samples for Flame Atomic-Emission and Absorption Spectroscopy
2. Boumans P. W. J. M., Theory of Spectrochemical Excitation (Plenum, New York, 1966), pp. 87–88.
Cited by
44 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献