Affiliation:
1. Battelle Memorial Institute, Pacific Northwest Laboratory, Richland, Washington 99352
Abstract
A total-burn spectrographic method has been applied to the determination of trace concentrations of silicon in plutonium. The sample is oxidized to plutonium dioxide, mixed with a germanium dioxide—graphite mixture containing tin internal standard, packed into cupped graphite electrodes, and burned to completion in a high-amperage dc arc. Plutonium spectral interference is avoided by using the less sensitive Si 2506.90-Å line. The lower limit of the determination is 25 ppm of silicon in plutonium. The precision of the method is estimated to be ±8% (coefficient of variation) at the 270-ppm level and ±14% at 40 ppm.
Subject
Spectroscopy,Instrumentation
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献