Affiliation:
1. Argonne National Laboratory, Argonne, Illinois
Abstract
An optical emission spectrographic method has been developed for the quantitative determination of molybdenum, ruthenium, rhodium, palladium, and zirconium in normal uranium metal in the range of 2.5%, 2.0%, 0.30%, 0.20%, and 2.5%, respectively. The precision of the spectrographic method for molybdenum is ±1.0%, ruthenium ±18%, rhodium ±17%, palladium ±21%, and zirconium ± 2.2% of the amounts present The comparison of spectrographic and chemical results on 24 “fissium” alloy samples for molybdenum, ruthenium, rhodium, palladium, and zirconium shows a standard relative deviation from the chemical results of ±3.9%, ±4.3%, ±7.2%, ±6.6% and ±7.5%, respectively. These samples were chosen at random from ingots by using routine sampling procedures
Subject
Spectroscopy,Instrumentation
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Precious Metals;Applied Atomic Spectroscopy;1978
2. EMISSION SPECTROSCOPIC METHODS;Analytical Chemistry of Molybdenum and Tungsten;1971
3. SPECTROGRAPHIC ANALYSES;Analytical Chemistry of Zirconium and Hafnium;1970
4. The Analysis of Uranium Alloys Using a Hollow Cathode;Applied Spectroscopy;1961-11