Affiliation:
1. Texas Eastman Division, Eastman Chemical Company, Longview, Texas 75607-7444 (J.B.T.); and Department of Chemistry, University of Arizona, Tucson, Arizona 85721-0041 (R.H.W., M.B.D.)
Abstract
A continuum source, multielement graphite furnace atomic absorption spectrometer was developed that employed a high-resolution echelle polychromator with charge-injection device (CID) detection. The detection system employed a new device, the CID 38, and camera control unit, the SCM5000E, providing unique new capabilities. A data acquisition and analysis scheme was developed for the prototype system, and the instrument demonstrated detection limits approaching that of commercial line source graphite furnace atomic absorption spectroscopy (GFAAS). However, the performance degrades in the ultraviolet due to decreasing source output.
Subject
Spectroscopy,Instrumentation
Cited by
9 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献