Affiliation:
1. Optical Sciences Division, Code 6505, Naval Research Laboratory, Washington, D.C. 20375-5000 (J.J., I.A.); and GEO-Centers, Inc., Fort Washington, Maryland 20744 (J.J.)
Abstract
This paper describes an analytical method for the quantitation of iron, cobalt, nickel, and copper at trace levels with graphite furnace-atomic absorption spectrometry (GF-AAS) without the use of any matrix modifier or preconcentration techniques. A solution of HfF4 (10%, w/v) in HF (24%) was used for this study. The detection limits for iron, cobalt, nickel, and copper were found to be 3, 1, 3, and 2 ppb, respectively. The average precision of measurements (% RSD) for all the elements was <10%.
Subject
Spectroscopy,Instrumentation
Cited by
8 articles.
订阅此论文施引文献